PEM Electrolyzer Test Bench


The GR-WETS-PEM-SC100 series electrolytic water hydrogen production cell testing equipment is designed to provide an accurate and reliable testing platform for electrolytic water hydrogen production cells. This testing device has undergone multiple upgrades, improvements, and optimizations based on the performance of existing products in the industry. The testing equipment consists of a deionized water circulation system, a nitrogen purging unit, a gas-water separation unit, a gas analysis pretreatment unit, a PLC acquisition and control unit, a human-machine operation unit, and a safety monitoring unit. It uses the company's self-developed system testing software, which can meet the testing requirements for polarization curves, hydrogen production energy efficiency, hydrogen quality testing, and product lifespan.

Product Details

Overview of PEM Electrolyzer Test Bench Product Power Levels

Serial Number

Number

Name

Model

1

1

(Desktop type; atmospheric pressure)

100W PEM Electrolyzer Test Bench

GR-WETS-PEM-SC100

2

500W PEM Electrolyzer Test Bench

GR-WETS-PEM-SC500

3

2

500W PEM Electrolyzer Test Bench

GR-WETS-PEM-SC500

4

1kW PEM Electrolyzer Test Bench

GR-WETS-PEM-S1K

5

5kW PEM Electrolyzer Test Bench

GR-WETS-PEM-S5K

6

30kW PEM Electrolyzer Test Bench

GR-WETS-PEM-S30K

7

100kW PEM Electrolyzer Test Bench

GR-WETS-PEM-S100K

8

500kW PEM Electrolyzer Test Bench

GR-WETS-PEM-S500K

9

1MW PEM Electrolyzer Test Bench

GR-WETS-PEM-S1M

10

5MW PEM Electrolyzer Test Bench

GR-WETS-PEM-S5M

11

3

(Multi-channel)

9-channel 200W PEM Electrolyzer Test Bench (Fully Independent)

GR-WETS-PEM-SC200X9D

12

4-channel 500W PEM Electrolyzer Test Bench (Fully Independent)

GR-WETS-PEM-SC500X4D

13

4-channel 500W PEM Electrolyzer Test Bench (Fully Shared)

GR-WETS-PEM-SC500X4G

 

Overview of PEM Electrolyzer Test Bench Products

1.1 Product Images

Serial Number

Type

Image

Image

1

Desktop PEM Electrolyzer Test Bench

2

PEM Electrolyzer Test Bench

3

Multi-channel PEM Electrolyzer Test Bench

 

 

1.2 Product Overview

The GR-WETS-PEM series electrolyzer test bench is designed to provide an accurate and reliable testing platform for electrolyzers. This test bench has undergone multiple upgrades and optimizations based on the performance of existing products in the industry. It consists of a deionized water circulation system, nitrogen purging unit, gas-water separation unit, gas analysis pre-treatment unit, PLC data acquisition and control unit, human-machine operation unit, and safety monitoring unit, utilizing the company's self-developed system testing software to meet the testing requirements for polarization curves, hydrogen production energy consumption efficiency, hydrogen quality testing, and product lifespan.

1.3 Product Functions

Nitrogen Purging: Equipped with automatic nitrogen purging function before and after testing, and during equipment failure/shutdown.

Electrolyte Control: Features a wide range of high-precision control for circulating water flow; conductivity monitoring function; liquid level monitoring and automatic water replenishment function.

Gas-Water Separation: Features an efficient gas-water separation design, including gas-liquid separation, condensation dehumidification, and drying devices.

Pressure Control: Equipped with automatic back pressure control for anode/cathode; features equal pressure and differential pressure control for both sides.

Flow Detection: Provides high-precision online measurement of gas flow.

Data Acquisition and Storage: Capable of real-time acquisition and monitoring of data such as current, voltage, temperature, pressure, and circulating water flow.

Safety Interlock and Protection: Features dual safety protection strategies (hardware and software), with three-level alarm parameters settable via upper computer software.

Performance Testing: Capable of testing electrolyzer polarization curves, chamber consistency, durability, sensitivity, and other testing functions.

Stability and Reliability: Capable of 24-hour unattended operation.

1.4 Product Advantages

High Control Stability: Steady-state temperature control ±1℃; pressure control ±0.5%F.S.; water flow control ±1%F.S.; conductivity control for purified water ≤0.1μS/cm.

High Measurement Accuracy: Temperature ±0.5℃; pressure ±0.5%F.S.; circulating water flow ±1%F.S.; gas flow ±0.8%RD+0.2%F.S.

High Integration Level: Highly integrated equipment, compact layout, aesthetically pleasing design, convenient operation, and complete interface functions.

Strong Data Processing Capability: Capable of automatically collecting, analyzing, and processing data, generating detailed test results including charts, data lists, etc.

High Equipment Safety: Complete safety protection and interlock functions, with three-level safety alarm limits settable via upper computer.

Strong Equipment Reliability: Equipment features various automatic condition testing functions, capable of 7×24 hours unattended operation.

Multiple Function Expansion: Can use nitrogen for pipeline pre-pressurization; can expand for quick sampling at the electrolyzer outlet for gas analysis; can integrate with electrochemical workstations for electrochemical testing; can expand to achieve electrolyzer polarization loss decomposition, EIS data analysis functions, etc.

1.5 Product Parameters

1.5.1 Desktop PEM Electrolyzer Test Bench

Model

GR-WETS-PEM-SC100

GR-WETS-PEM-SC500

Rated Power

100W

500W

Electrolyzer Type

PEM Electrolyzer

System Module

Technical Specifications

Specific Parameters

Deionized Water Circulation Module

Medium

Deionized Water

Circulation Mode

Anode Deionized Water Circulation

Conductivity

Features online conductivity detection; detection range: 0~20μS/cm; accuracy: ±0.5%F.S.

Purification Function

Features online deionization function, conductivity control ≤0.1μS/cm.

Water flow rate

0.05~0.5L/min

Temperature control

RT~90℃; Steady-state temperature control accuracy: ±1℃; Heating rate: 1~4℃/min

Automatic water replenishment

Equipped

Hydrogen/Oxygen system module

Hydrogen flow rate

0~0.5SLPM

0~2.5SLPM

Back pressure range

Atmospheric pressure

Dehydration function

Equipped with gas-liquid separation, condensation dehumidification, drying

Nitrogen purging

Purging function

Equipped with automatic/manual nitrogen purging function before/after testing, customizable purging time

Gas analysis

Hydrogen in oxygen

Range: 0~4%VOL; Accuracy: ±2%F.S. (Calibration available for range 0~5%VOL)

Power module

Working mode

Constant pressure, constant current, constant power, programmable automatic operation

Voltage

0~20V

0~16V

Current

0~100A

0~250A

Voltage inspection

Voltage inspection

Equipped with individual voltage inspection function; Number of channels: customizable; Range: ±3V; Accuracy: ±1mV; Alarm limit can be set

Whole system

Controller

PLC

Host computer

Optional C# or Labview

Communication interface

LAN

Power supply

AC220V/50Hz

Overall structure

Desktop type

Optional functions

Fixtures; Oxygen flow detection; Hydrogen in oxygen detection; Dew point detection; Electrochemical workstation; Electrochemical testing; Polarization loss decomposition and EIS data analysis functions

Note: Upgrades or downgrades can be made according to requirements.

1.5.2 PEM electrolyzer test bench

Model

GR-WETS-PEM-SC500

GR-WETS-PEM-S1K

GR-WETS-PEM-S5K

GR-WETS-PEM-S30K

GR-WETS-PEM-S100K

GR-WETS-PEM-S500K

GR-WETS-PEM-S1M

GR-WETS-PEM-S5M

Rated Power

500W

1kW

5kW

30kW

100kW

500kW

1MW

5MW

Electrolyzer Type

PEM Electrolyzer

System Module

Technical Specifications

Specific Parameters

Deionized Water Circulation Module

Medium

Deionized Water

Circulation Mode

Anode Deionized Water Circulation

Conductivity

Features online conductivity detection; detection range: 0~20μS/cm; accuracy: ±0.5%F.S.

Purification Function

Features online deionization function, conductivity control ≤0.1μS/cm.

Water flow rate

1L/min

2L/min

10L/min

50L/min

180L/min

500L/min

800L/min

4000L/min

Temperature control

RT~90℃; Steady-state temperature control accuracy: ±1℃; Heating rate: 1~4℃/min

Automatic water replenishment

Equipped

Hydrogen/Oxygen system module

Hydrogen flow rate

2.5SLPM

5SLPM

25SLPM

150SLPM

500SLPM

2500SLPM

5000SLPM

25000SLPM

Back pressure range

Maximum 0~5MPa; Control accuracy: ±0.5%F.S.; Equipped with gas equalization and differential pressure control functions on both sides

Dehydration function

Equipped with gas-liquid separation, condensation dehumidification, drying

Nitrogen purging

Purging function

Equipped with automatic/manual nitrogen purging function before/after testing, customizable purging time

Gas analysis

Hydrogen in oxygen

Range: 0~4%VOL; Accuracy: ±2%F.S. (Calibration available for range 0~5%VOL)

Power module

Working mode

Constant pressure, constant current, constant power, programmable automatic operation

Voltage

Custom services provided based on actual needs

Current

Custom services provided based on actual needs

Voltage inspection

Voltage inspection

Equipped with individual voltage inspection function; Number of channels: customizable; Range: ±3V; Accuracy: ±1mV; Alarm limit can be set

Whole system

Controller

PLC

Host computer

Optional C# or Labview

Communication interface

LAN

Power supply

AC220V/50Hz or AC380V/50Hz

Overall structure

Rack type

Cabinet/container/skid type

Optional functions

Oxygen flow detection; Hydrogen in oxygen detection; Dew point detection; Nitrogen pre-pressurization; Sampling gas analysis; Electrochemical testing; Polarization loss decomposition and EIS data analysis functions

Note: Upgrades or downgrades can be made according to requirements.

1.5.3 Multi-channel PEM electrolyzer test bench

Model

GR-WETS-PEM-SC200X9D

GR-WETS-PEM-SC500X4D

GR-WETS-PEM-SC500X4G

Single channel rated power

200W

500W

Electrolyzer Type

PEM Electrolyzer

Channel form

9 independent channels

4 independent channels

4 shared channels

System Module

Technical Specifications

Specific Parameters

Deionized Water Circulation Module

Medium

Deionized Water

Circulation Mode

Anode deionized water circulation, each channel is independent

Conductivity

Each channel is equipped with online conductivity detection function

Detection range: 0~20μS/cm; Accuracy: ±0.5%F.S.

Shared 1 set of online conductivity detection function

Detection range: 0~20μS/cm; Accuracy: ±0.5%F.S.

Purification Function

Each channel is equipped with online deionization function

Conductivity control ≤0.1μS/cm

Shared 1 set of online deionization function

Conductivity control ≤0.1μS/cm

Water flow rate

0.05~0.5L/min

Temperature control

RT~90℃; Steady-state temperature control accuracy: ±1℃; Heating rate: 1~4℃/min

Automatic water replenishment

Equipped

Hydrogen/Oxygen system module

Hydrogen flow rate

Single channel 0~1SLPM

Single channel 0~2.5SLPM

0~10SLPM

Back pressure range

9 independent channels back pressure, maximum 0~5MPa

4 independent channels back pressure, maximum 0~5MPa

4 shared channels back pressure, maximum 0~5MPa

Dehydration function

All 9 channels are equipped with gas-liquid separation, condensation dehumidification, drying

All 4 channels are equipped with gas-liquid separation, condensation dehumidification, drying

Shared 1 set of gas-liquid separation, condensation dehumidification, drying

Nitrogen purging

Purging function

Equipped with automatic/manual nitrogen purging function before/after testing, customizable purging time

Gas analysis

Hydrogen in oxygen

Range: 0~4%VOL; Accuracy: ±2%F.S. (Calibration available for range 0~5%VOL)

Single channel power module

Working mode

Constant pressure, constant current, constant power, programmable automatic operation

Voltage

0~10V

0~16V

Current

0~200A

0~250A

Voltage inspection

Voltage inspection

Equipped with individual voltage inspection function; Number of channels: customizable; Range: ±3V; Accuracy: ±1mV; Alarm limit can be set

Whole system

Controller

PLC

Host computer

Optional C# or Labview

Communication interface

LAN

Power supply

AC220V/50Hz or AC380V/50Hz

Overall structure

Rack type

Optional functions

Fixtures; Independent detection/switching inspection of oxygen flow; Switching inspection of hydrogen in oxygen; Independent detection/switching inspection of oxygen in hydrogen; Dew point detection; Nitrogen pre-pressurization; Electrochemical testing; Polarization loss decomposition and EIS data analysis functions

Note: Upgrades or downgrades can be made according to requirements.

 